Medtronic Ares Antibiotic-Impregnated Catheter, REF: 93092, Sterile, Rx Only. Used in the treatment of hydrocephalus as a component of a shunt system when draining or shunting of cerebrospinal fluid (CSF) is indicated.
Z-0739-2021
Product summary
- Event
- Event 86988
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 1,926 units
- Official record key
device-enforcement:Z-0739-2021
Official wording
Reason: There is a potential for a defect on the seal of the outer pouch of catheters. The defect may compromise the sterility of the pouch contents, which may increase the risk of post-operative infection, requiring further medical intervention.
Code information: Lot numbers: 0010258435, 0010281390, 0010312360, 0010353505, 0010384900, 0010273889, 0010281391, 0010312361, 0010353507, 0010384901, 0010273890, 0010281392, 0010312362, 0010369366, 0010384903, 0010273891, 0010289835, 0010316959, 0010369367, 0010393220, 0010273893, 0010289836, 0010316960, 0010376784, 0010393221, 0010278418, 0010297678, 0010316961, 0010376785, 0010278427, 0010305929, 0010316963, 0010376786, 0010281384, 0010305937, 0010336620, 0010384897
Distribution pattern: Worldwide Distribution. US nationwide, Albania, Argentina, Armenia, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Finland, Germany, Greece, Guam, Hong Kong, Hungary, India, Islamic Republic of Iran, Ireland, Israel, Italy, Latvia, Macao, Malaysia, Mauritius, Republic of Moldova, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Poland, Portugal, Qatar, Saudi Arabia, Serbia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, and Vietnam
Derived failure modes
-
Sterility assurance
seal of the outer pouch of catheters. The defect may compromise the sterility