Device product
Microsurgical Operating Room Tables;7300BX (18-351-70).
Z-0567-2021
Product summary
- Event
- Event 85442
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 10
- Official record key
device-enforcement:Z-0567-2021
Official wording
Reason: Unintended tilting of the Tabletop MST 7300BX and incorrect operation due to infrared operating signal errors of Tabletops MST 7300B and MST 7300BX.
Code information: Tilt Movement-Leaking Valve (7300BX-18-351-70) serial numbers: MUC-20163-0001, MUC-2016Z-0002, MUC-2016Z-0003, MUC-20177-0005, MUC-20177-0007, MUC-20177-0008, MUC-20177-0009, MUC-20189-0017 & MUC-2018Y-0018. Unintended movement due to infrared signal errors (7300BX-18-351-70) serial numbers: MUC-20163-0001, MUC-2016Z-0002, MUC-2016Z-0003, MUC-20177-0005, MUC-20177-0007, MUC-20177-0008, MUC-20177-0009, MUC-20189-0017, MUC-2018Y-0018 & MUC-2018Y-0019.
Distribution pattern: US: AZ, CA, FL, MO, NC, OH, PA, TX, and WI. OUS: Greece, Israel, Japan, and Ukraine.
Derived failure modes
-
Unknown
Unintended tilting of the Tabletop MST 7300BX and incorrect operation due to infrared operating signal errors of Tabletops MST 7300B and MST 7300BX.