Device product
Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-32-XX, Acumatch GXL 0 Degree Liner, 32mm; and Novation GXL Liner, Neutral, 32mm. Orthopedic hip implant component.
Z-2117-2021
Product summary
- Firm
- Exactech, Inc.
- Event
- Event 88126
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 89,050 total
- Official record key
device-enforcement:Z-2117-2021
Official wording
Reason: Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
Code information: All serial numbers. UDI-DI numbers: 10885862022097, 10885862022103, 10885862022110, 10885862022127, 10885862207074, 10885862022165, 10885862022172, 10885862022189, 10885862022196
Distribution pattern: Distribution US nationwide.
Derived failure modes
-
Unknown
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.