Device product

Raystation treatment planning system for radiation therapy, Model Numbers 9.0.0, 9.0.1, 9.1.0, 9.2.0, 10.0.0, 10.0.1, 10.0.2, 10.1.0, 10.1.1, 11.0.0, 11.0.1, 11.0.3, 11.0.4, 12.0.0, 12.1.0, 12.1.1, 12.0.3, 12.1.2, 12.0.4, 12.1.3, 13.0.0, 13.1.0. Software Version Numbers 9A, 9B, 10A, 10B, 11A, 11B and 12A including service packs

Z-1281-2023

February 17, 2023

Class II

Product summary

Firm
RAYSEARCH LABORATORIES AB
Event
Event 91867
Status
Ongoing
Classification
Class II
Quantity
10770 systems
Official record key
device-enforcement:Z-1281-2023

Official wording

Reason: It is possible to set a non-zero collimator angle in the Virtual Simulation module. The angle will be correct in the user interface but if the plan is DICOM exported by using the VSIM Export button, the angle will not be correct in the exported RT plan. In the exported RT plan, the DICOM attribute Beam Limiting Device Angle (300A,0120) is set to zero, regardless of the selected collimator angle.

Code information: UDI-DI: 0735000201067920221007;

Distribution pattern: US Nationwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    It is possible to set a non-zero collimator angle in the Virtual Simulation module. The angle will be correct in the user interface but if the plan is DICOM exported by using the VSIM Export button, the angle will not be correct in the exported RT plan. In the exported RT plan, the DICOM attribute Beam Limiting Device Angle (300A,0120) is set to zero, regardless of the selected collimator angle.