Device product
DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B
Z-1283-2023
Product summary
- Event
- Event 91760
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 27,200 devices
- Official record key
device-enforcement:Z-1283-2023
Official wording
Reason: The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum.
Code information: GTIN 50749756377046, Lot 58447585 exp. 12/22/2027
Distribution pattern: US Nationwide distribution in the states of Arizona, Texas, Virginia, Pennsylvania, California, Missouri, North Carolina, Illinois, Indiana, Ohio, Arkansas.
Derived failure modes
-
Unknown
The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum.