Lyra SARS-CoV-2 Assay (M120) in vitro diagnostic test reagent

Z-1928-2021

April 27, 2021

Class I

Product summary

Firm
Quidel Corporation
Event
Event 87996
Status
Terminated
Classification
Class I
Quantity
29,787 kits (96 results/kit)
Official record key
device-enforcement:Z-1928-2021

Official wording

Reason: In rare situations patients infected with high viral load of SARS CoV 2 may have samples that generate Ct values less than or equal to (d) 5 when certain amplification and detection platforms (thermocyclers) are used. These specimens would be falsely interpreted as negative if target amplification occurs prior to a cycle threshold (Ct) value of 5. There have been no formal complaints of False Negative results with Ct values <5 from Lyra SARS-CoV-2 Assay users.

Code information: Lot Number: 031620A 031620B 031620C 032320 032420 032720 032820A 032820B 040320 040720 040920 041020 174992 175429 175501 175502 175503 176001 176002 176366 176367 176368 178984 178985 180331 180332 180673 180674 180675 182594 184273 185535 185822 186470 186472 187062 187173 187822 189232 189942 190786 193074 193977

Distribution pattern: Worldwide distribution - US Nationwide distribution in the states of AL, CA, CT, DC, FL, ID, IN, MA, MI, MO, NM, NY, OH, SC, TN, TX, WA and the countries of Canada, Saudi Arabia, United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    In rare situations patients infected with high viral load of SARS CoV 2 may have samples that generate Ct values less than or equal to (d) 5 when certain amplification and detection platforms (thermocyclers) are used. These specimens would be falsely interpreted as negative if target amplification occurs prior to a cycle threshold (Ct) value of 5. There have been no formal complaints of False Negative results with Ct values <5 from Lyra SARS-CoV-2 Assay users.