Device product
UroLift Visual Obturator, 20Ch-The visual obturator is used with the UroLift¿ system. Instrument for insertion of a cystoscope shaft into the bladder. Model: UL-VO
Z-2364-2023
Product summary
- Event
- Event 92700
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 98 units
- Official record key
device-enforcement:Z-2364-2023
Official wording
Reason: There is an improperly performed weld between the cone of the sheath lock and tube of the obturator; mechanical load capacity is impaired, and fracture of the welded joint may occur. Contamination can occur in this capillary gap, which is not covered by the reprocessing validations performed.
Code information: UDI-DI: 04250480106617 Lot Number: 141980
Distribution pattern: Nationwide Foreign: Japan, South Korea
Derived failure modes
-
Unknown
There is an improperly performed weld between the cone of the sheath lock and tube of the obturator; mechanical load capacity is impaired, and fracture of the welded joint may occur. Contamination can occur in this capillary gap, which is not covered by the reprocessing validations performed.