Skip to content
Recall Observatory FDA recall evidence

Device product

Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 104-36-XX, MCS +5GXL Liner 5/15 DEG and MCS GXL Liner 5/15 DEG. Orthopedic hip implant component.

Z-2114-2021

June 29, 2021

Class II

Product summary

Firm
Exactech, Inc.
Event
Event 88126
Status
Ongoing
Classification
Class II
Quantity
89,050 total
Official record key
device-enforcement:Z-2114-2021

Official wording

Reason: Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.

Code information: All serial numbers. UDI-DI numbers: 10885862013682, 10885862013699, 10885862013705, 10885862013712, 10885862013729

Distribution pattern: Distribution US nationwide.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.