Skip to content
Recall Observatory FDA recall evidence

Device product

VITROS Immunodiagnostic Products SARS-CoV-2 Antigen Reagent- In Vitro Diagnostic. The VITROS Immunodiagnostic Products SARS-CoV-2 Antigen Reagent Pack when used in combination with the VITROS Immunodiagnostic Products SARS-CoV-2 Antigen Calibrator is a chemiluminescent immunoassay intended for the qualitative detection of SARS-CoV-2 nucleocapsid protein antigens in nasopharyngeal (NP) specimens. . For emergency authorization use only (EAU). Intended for Use in the United States Product Code: 6199949.

Z-1266-2021

February 12, 2021

Class II

Product summary

Firm
Ortho-Clinical Diagnostics, Inc.
Event
Event 87356
Status
Terminated
Classification
Class II
Quantity
2782 units
Official record key
device-enforcement:Z-1266-2021

Official wording

Reason: VITROS Immunodiagnostic Products SARS-CoV-2 Antigen False Reactive Results Associated When Using Remel M4RT Viral Transport Media

Code information: Any lot used with Remel M4RT VTM UDI: 10758750033584

Distribution pattern: Worldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Canada L3R 4G5, Chile, Colombia, Denmark, France, Germany, India, Italy, Japan, Portugal, Russia, Singapore 609917, Spain, United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0