Device product
VADER pedicle system, torque wrench, Catalog Number 42-702
Z-0700-2024
Product summary
- Firm
- Icotec Ag
- Event
- Event 93577
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 79 devices (US only)
- Official record key
device-enforcement:Z-0700-2024
Official wording
Reason: icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was not functioning correctly. Due to the malfunction, the nut screws may not have been tightened sufficiently.
Code information: 19/01-001, 19/01-013, 19/01-014, 19/01-015, 190830-10, 190830-23, 190830-42, 190830-44, 190830-45, 190830-50, 190830-51, 190830-52, 200131-54, 200131-56, 200131-62, 200131-64, 200131-68, 200131-69, 200131-70, 200131-72, 200131-73, 200131-74, 200221-040R1, 200424-01, 200424-03, 200424-04, 200424-05, 200424-06, 200424-10, 200424-11, 200424-12, 200424-13, 200424-14, 200424-15, 200424-16, 200424-17, 200424-20, 200424-21, 200424-22, 200424-24, 200424-25, 200424-27, 200424-29, 200424-32, 200424-34, 200424-35, 200605-051R1, 200605-052R1, 201002-073R1, 210915-005, 210915-009, 210915-010, 210915-011, 210915-012, 220610-001, 220610-003, 220610-008, 220610-019, 220610-020, 220610-021, 220610-025, 220610-026, 220610-030, 230113-003, 230113-005, 230113-007, 230113-008, 230113-009, 230113-010, 230113-011, 230113-012, 230113-013, 230113-014, 230224-029, 230224-034, 230426-038, 230426-039, 230426-042, 230426-043
Distribution pattern: US: AZ. MI, OH, CA, CO, TX, NY, NC, DE, PA
Derived failure modes
-
Unknown
icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was not functioning correctly. Due to the malfunction, the nut screws may not have been tightened sufficiently.