Device product
Medtronic Reveal LINQ LNQ11 / PA96000
Z-2102-2021
Product summary
- Event
- Event 88175
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 666210 devices
- Official record key
device-enforcement:Z-2102-2021
Official wording
Reason: Reveal LINQ with TruRhythm ICMs, that undergo a partial electrical reset appear to be programmed "ON", but are no longer able to detect and report Brady and Pause events to clinicians.
Code information: All serial numbers distributed on or after February 23, 2017 GTIN: 643169845763, 643169845770, 643169845787, 643169845794, 643169845800, 643169845817, 643169845824, 00643169845831, 00643169845848, 00643169845855, 00643169845862, 00643169845879, 00643169845893, 00763000188382, 00763000188399, 00763000188405, 00763000188412, 00763000188429, 00763000188436, 00763000188443, 00763000188450, 00763000188467, 643169845688, 643169845695, 643169845701, 643169845718, 643169845732, 643169845749, 643169845756
Distribution pattern: Worldwide Distribution
Derived failure modes
-
Unknown
Reveal LINQ with TruRhythm ICMs, that undergo a partial electrical reset appear to be programmed "ON", but are no longer able to detect and report Brady and Pause events to clinicians.