Device product

Medtronic Reveal LINQ LNQ11 / PA96000

Z-2102-2021

June 01, 2021

Class II

Product summary

Firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Event
Event 88175
Status
Ongoing
Classification
Class II
Quantity
666210 devices
Official record key
device-enforcement:Z-2102-2021

Official wording

Reason: Reveal LINQ with TruRhythm ICMs, that undergo a partial electrical reset appear to be programmed "ON", but are no longer able to detect and report Brady and Pause events to clinicians.

Code information: All serial numbers distributed on or after February 23, 2017 GTIN: 643169845763, 643169845770, 643169845787, 643169845794, 643169845800, 643169845817, 643169845824, 00643169845831, 00643169845848, 00643169845855, 00643169845862, 00643169845879, 00643169845893, 00763000188382, 00763000188399, 00763000188405, 00763000188412, 00763000188429, 00763000188436, 00763000188443, 00763000188450, 00763000188467, 643169845688, 643169845695, 643169845701, 643169845718, 643169845732, 643169845749, 643169845756

Distribution pattern: Worldwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Reveal LINQ with TruRhythm ICMs, that undergo a partial electrical reset appear to be programmed "ON", but are no longer able to detect and report Brady and Pause events to clinicians.