Device product
Simpleware ScanIP medical software. For transfer of imaging information from a medical scanner to an output file, and for simulating/evaluating surgical treatment options.
Z-2254-2021
Product summary
- Event
- Event 88207
- Status
- Terminated
- Classification
- Class II
- Quantity
- 6 software's
- Official record key
device-enforcement:Z-2254-2021
Official wording
Reason: A issue (bug) has been identified with the interface and image software which could result in anatomical orientation tags/labels being displayed incorrectly. This could result in operator misunderstanding and errors during resampling or re-alignment of image data
Code information: Affected version/UDI: Simpleware ScanIP Medical; Simpleware ScanIP O-2018.12: 00863520000338, Simpleware ScanIP Medical; Simpleware ScanIP P-2019.09: 00863520000345, Simpleware ScanIP Medical; Simpleware ScanIP Q-2020.03: 00863520000369, Simpleware ScanIP Medical; Simpleware ScanIP Q-2020.06: 00863520000352, Simpleware ScanIP Medical; Simpleware ScanIP R-2020.09: 00863520000376, Simpleware ScanIP Medical; Simpleware ScanIP R-2021.03: 00863520000383
Distribution pattern: US: Nationwide Distribution in the states of CA, FL and countries of : France, United Kingdom, Germany, Australia
Derived failure modes
-
Device software or design
issue (bug) has been identified with the interface and image software