Skip to content
Recall Observatory FDA recall evidence

Device product

Naturalyte Liquid Acid concentrate drums, Product Code 13-3251-9 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

Z-0840-2021

November 23, 2020

Class II

Product summary

Firm
Fresenius Medical Care Holdings, Inc.
Event
Event 86939
Status
Ongoing
Classification
Class II
Quantity
2 drums
Official record key
device-enforcement:Z-0840-2021

Official wording

Reason: The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.

Code information: Lot Numbers: 20LXAC066 20LXAC089 Additional Affected lots as of 3/17/21: 20LXAC066 20LXAC073 20LXAC074 20LXAC024 20LXAC023 20LXAC072 20LXAC085 20LXAC089

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.