Device product
4500MD Triple Quad Mass Spectrometer, Part Number 5032522
Z-0038-2024
Product summary
- Firm
- AB Sciex, LLC
- Event
- Event 92810
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 38
- Official record key
device-enforcement:Z-0038-2024
Official wording
Reason: In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
Code information: UDI-DI: 00842365102049; Serial Numbers: EJ230051806PL, EJ230061807PL, EJ230071808, EJ230101810, EJ230111811, EJ230121811, EJ230141811, EJ230151812, EJ230161812, EJ230171901, EJ230181901, EJ230191901, EJ230211905, EJ230221908, EJ230231908, EJ230251909, EJ230271910, EJ230281910, EJ230311911, EJ230331912, EJ230362003, EJ230372003, EJ230392004, EJ230412009, EJ230422011, EJ230442012, EJ230452012, EJ230472101, EJ230482103, EJ230512105, EJ240012110, EJ240022111, EJ240052111, EJ240082207, EJ240092209, EJ240142307, EJ240132305, EJ240062201
Distribution pattern: US Nationwide.
Derived failure modes
-
Unknown
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.