Device product

Ingenia 1.5T Evoluton- Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 781315

Z-2341-2021

July 30, 2021

Class II

Product summary

Firm
Philips North America Llc
Event
Event 88461
Status
Terminated
Classification
Class II
Quantity
2 units US; 16 units OUS
Official record key
device-enforcement:Z-2341-2021

Official wording

Reason: Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury

Code information: S/N and UDI: US: 61044 (01)00884838099715(21)61044; 61047 (01)00884838099715(21)61047. OUS: 61028 (01)00884838099715 61008 (01)00884838099715 61025 (01)00884838099715 61026 (01)00884838099715 61034 (01)00884838099715 61035 (01)00884838099715 61042 (01)00884838099715 61046 (01)00884838099715 61040 (01)00884838099715 61041 (01)00884838099715 61018 (01)00884838099715 61017 (01)00884838099715 61032 (01)00884838099715 61043 (01)00884838099715 61033 (01)00884838099715 61064 (01)00884838099715

Distribution pattern: Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, MD, MD, NH, NY, OH, TN, TX, VA and the countries of Canada, Mexico, Australia, Austria, Chile, China, Czech Republic, Denmark, France, Germany, India, Indonesia, Italy, Japan, Korea, Republic of Lebanon, Netherlands, Romania, Russian Federation, Saudi Arabia, Spain, Turkey, Ukraine, United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury