Device product
Veran : Percutaneous Models: INS-5620 SPiN Perc Biopsy Needle Guide Kit - 20cm, 5/Bx (includes INS-5028)" INS-5611 SPiN Perc¿ Localization Needle - 2cm, 5/Bx INS-5610 SPiN Perc¿ Localization Needle - 1cm, 5/Bx INS-5600 SPiN Perc¿ Biopsy Needle Guide Kit, 5/Bx (includes INS-5029)
Z-2661-2023
Product summary
- Event
- Event 92893
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 9961 units
- Official record key
device-enforcement:Z-2661-2023
Official wording
Reason: Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment
Code information: Model-UDI: INS-5620 SPiN Perc Biopsy Needle Guide Kit - 20cm, 5/Bx (includes INS-5028)" 00815686021122; INS-5611 SPiN Perc¿ Localization Needle - 2cm, 5/Bx 00815686021146; INS-5610 SPiN Perc¿ Localization Needle - 1cm, 5/Bx 00815686021139; INS-5600 SPiN Perc¿ Biopsy Needle Guide Kit, 5/Bx (includes INS-5029) 00815686020736. All serial numbers and all lot numbers
Distribution pattern: Worldwide distribution - US Nationwide and the countries of CN, DE, HK, IT, KR, SG, TH, TW.
Derived failure modes
-
Unknown
Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment