Device product
Citrine QTRAP Mass Spectrometer, Part Number 5063685
Z-0039-2024
Product summary
- Firm
- AB Sciex, LLC
- Event
- Event 92810
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 10
- Official record key
device-enforcement:Z-0039-2024
Official wording
Reason: In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
Code information: UDI-DI: 00842365100540; Serial Numbers: DD250051901, DD250172009, DD250182009, DD230001804, DD250111910, DD250132004, DD250091909, DD250101910, DD250192010, CN10011707PT
Distribution pattern: US Nationwide.
Derived failure modes
-
Unknown
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.