Skip to content
Recall Observatory FDA recall evidence

Device product

Route 92 Medical Full Length 070 Access System, REF 7002, Rxonly, SterileEO. Used for Introduction of interventional devices into the neurovasculature.

Z-1610-2024

March 08, 2024

Class I

Product summary

Firm
ROUTE 92 MEDICAL INC
Event
Event 94270
Status
Ongoing
Classification
Class I
Quantity
10 units
Official record key
device-enforcement:Z-1610-2024

Official wording

Reason: Their is the potential for distal tip separation at the proximal marker band of the delivery catheters.

Code information: UDI: 00853799007473/ Lot # 23051901

Distribution pattern: US: AL AZ CA DE FL GA IA IL IN KY LA MA MD MI MN MO NC ND NJ NY OH OR PA RI SD TN TX UT VA WV OUS: New Zealand

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0