Device product
Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 728333 - Spectral CT:
Z-1986-2021
Product summary
- Event
- Event 87974
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 1,301 Monitor Bracket Assemblies total
- Official record key
device-enforcement:Z-1986-2021
Official wording
Reason: Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
Code information: Model/ Serial Number/UDI: N/A 10016 (01)00884838101111(21)10016 10018 (01)00884838101111(21)10018
Distribution pattern: Worldwide distribution - US Nationwide and the countries of Canada, Mexico, AE, AR, AT, AU, BD, BE, BR, CH, CL, CN, CO, CU, CY, CZ, DE, DK, DZ, EE, EG, ES, FI, FR, GB, GE, GI, GL, GR, HK, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KG, KH, KR, LA, LB, LK, LT, LY, MM, MN, MU, MV, MY, NI, NL, NO, NZ, OM, PA, PF, PH, PK, PL, PS, PT, QA, RE, RU, RW, SA, SE, SG , SI, SK, TH, TN, TR, UZ, VN, XK, YE, ZA, ZW.
Derived failure modes
-
Unknown
Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.