Device product
Speed Stitch Needle Cassette
Z-2999-2020
Product summary
- Event
- Event 86290
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 6,366 needles
- Official record key
device-enforcement:Z-2999-2020
Official wording
Reason: Due to a supplier error where a portion of the raw material was inadvertently mixed. As a result, some needles may be made of commercially pure titanium as opposed to 304 stainless steel which does not meet the specification and could potentially break during use.
Code information: Model: OM-8850 Lot Numbers: 2018434, 2018934, 2019558, 2019559, 2019929, 2019930, 2020931, 2022111, 2022641, 2030008, 2032212, 2033815, 2037604, 2036941, 2038701, 2039699, 2040500, 2042295, 2044957, 2047483, 2048575, 2051239 & 2032526
Distribution pattern: Domestic Distribution: AK, AL, AR, AZ, CA, CO, CT,DB,FL,GA, HI, IA, ID, IL, IN, KS, KY, MA,MD,ME,MI,MN,MO,MS,MT, NC,ND, NE, NJ, NSW, NV,NY, OH,OK, OR, PA,SC, TN,TX, UT, VA, WA, WI, WV. International Distribution: AE, AU, CH, DE, FI, FR, GB, HK, IE, PA, PR, PT, RU, SG, and ZA
Derived failure modes
-
Potency or specification failure
does not meet the specification