United Orthopedics USTAR II Knee System USTAR II Knee System Segment Part, RHS, 25 mm Length, Catalog # 2915-1025

Z-0110-2022

August 27, 2021

Class II

Product summary

Firm
Uoc Usa Inc
Event
Event 88692
Status
Terminated
Classification
Class II
Quantity
3 devices
Official record key
device-enforcement:Z-0110-2022

Official wording

Reason: Due to product not meeting specification for concentricity and failing to meet requirements of dynamic fatigue testing.

Code information: Catalog Number: 2915-1025 UDI Code: 04719872148197 Lot Code: 19E272H

Distribution pattern: U.S. Nationwide: CA, CO, and NV No O.U.S.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0