Device product

uCT 760 System, Computed Tomography X-ray System

Z-2522-2020

May 08, 2020

Class II

Product summary

Firm
Shanghai United Imaging Healthcare Co., Ltd.
Event
Event 85804
Status
Terminated
Classification
Class II
Quantity
5 units
Official record key
device-enforcement:Z-2522-2020

Official wording

Reason: Two issues were identified with the computed tomography x-ray system including an intermittent issue which may affect image data leading to artifacts; and intermittent scout scanning interruption due to occasional angle signal drift. This may result in the need for rescanning of the patient with the patient receiving an additional dose of radiation and possible use of additional contrast media.

Code information: Model uCT 760; Serial Numbers: 600229, 600230, 600232, 600271 with software version R001.11.0.729741-Re-1100, and serial number 600282 with software version R001.11.0.785930-Re-1150.

Distribution pattern: Distributed to two customers. Both customers located in Houston, Texas.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Two issues were identified with the computed tomography x-ray system including an intermittent issue which may affect image data leading to artifacts; and intermittent scout scanning interruption due to occasional angle signal drift. This may result in the need for rescanning of the patient with the patient receiving an additional dose of radiation and possible use of additional contrast media.