Device product
OT1000 Series Orthopedic Surgical Tables
Z-1149-2021
Product summary
- Event
- Event 87230
- Status
- Terminated
- Classification
- Class II
- Quantity
- 272 tables total (Domestic: 227 tables, Foreign 45 tables)
- Official record key
device-enforcement:Z-1149-2021
Official wording
Reason: On January 7, 2021, STERIS identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation occurred during a patient procedure.
Code information: Model No.: OT1000, OT1100, OT1200
Distribution pattern: Domestic: AL, AK, AR, AZ, CA, CO, CT, DC, FL, GA, ID, IL IN, IW, KS, KY, LA, ME, MD, MI, MN, MO, NE, NV, NJ, NY, NC, ND, OH, OK, OR, PA, PR, SC, TX, UT, VT, VA, WA, WV, WI. Foreign: Canada, South Korea, Panama, Qatar, Thailand.
Derived failure modes
-
Unknown
On January 7, 2021, STERIS identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation occurred during a patient procedure.