DABRA Laser (RA-308 Excimer Laser) Intended Use: For use in ablating a channel in occlusive peripheral vascular disease. Indications: For crossing chronic total occlusions in patients with symptomatic infrainguinal lower extremity vascular disease.

Z-0569-2021

October 16, 2020

Class II

Product summary

Firm
Ra Medical Systems, Inc.
Event
Event 86692
Status
Terminated
Classification
Class II
Quantity
36 devices
Official record key
device-enforcement:Z-0569-2021

Official wording

Reason: Due to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on the footswitch, it may fail in the 'on' position and allow the laser to continue to lase.

Code information: Model Number RA-308 Product/Catalog Number: 1200-0001-01 Serial Numbers: RA00027, RA00046, RA00049, RA00050, RA00054, RA00056, RA00057, RA00058, RA00064, RA00069, RA00073, RA00081, RA00083, RA00084, RA00085, RA00089, RA00097, RA00098, RA00114, RA00115, RA00120, RA00121, RA00125, RA00127, RA00135, RA00138, RA00148, RA00153, RA00154, RA00160, RA00164, RA00165, RA00167, RA00170, RA00175, and RA00177.

Distribution pattern: US Distribution to states of: AL, AZ, CA, FL, IL, IN, KY, LA, MD, NC, NV, NY, OK, OR, TX, and WA; and to OUS country: Japan.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Due to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on the footswitch, it may fail in the 'on' position and allow the laser to continue to lase.