DABRA Laser (RA-308 Excimer Laser) Intended Use: For use in ablating a channel in occlusive peripheral vascular disease. Indications: For crossing chronic total occlusions in patients with symptomatic infrainguinal lower extremity vascular disease.
Z-0569-2021
Product summary
- Event
- Event 86692
- Status
- Terminated
- Classification
- Class II
- Quantity
- 36 devices
- Official record key
device-enforcement:Z-0569-2021
Official wording
Reason: Due to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on the footswitch, it may fail in the 'on' position and allow the laser to continue to lase.
Code information: Model Number RA-308 Product/Catalog Number: 1200-0001-01 Serial Numbers: RA00027, RA00046, RA00049, RA00050, RA00054, RA00056, RA00057, RA00058, RA00064, RA00069, RA00073, RA00081, RA00083, RA00084, RA00085, RA00089, RA00097, RA00098, RA00114, RA00115, RA00120, RA00121, RA00125, RA00127, RA00135, RA00138, RA00148, RA00153, RA00154, RA00160, RA00164, RA00165, RA00167, RA00170, RA00175, and RA00177.
Distribution pattern: US Distribution to states of: AL, AZ, CA, FL, IL, IN, KY, LA, MD, NC, NV, NY, OK, OR, TX, and WA; and to OUS country: Japan.
Derived failure modes
-
Unknown
Due to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on the footswitch, it may fail in the 'on' position and allow the laser to continue to lase.