Device product
Veradius Unity, Model #718132
Z-0649-2022
Product summary
- Event
- Event 89437
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 12 US; 44 ROW
- Official record key
device-enforcement:Z-0649-2022
Official wording
Reason: Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not connected to the system.
Code information: Wireless Foot switch 3P : 459801238191 Wireless Foot switch 3P: 459801238231
Distribution pattern: Domestic distribution to AZ CA FL GA HI IL KS KY MA MD ME MI MN MS NC NE NJ NV NY OH PA TN TX VA WA WV. Worldwide foreign distribution.
Derived failure modes
-
Device software or design
firmware issue