Skip to content
Recall Observatory FDA recall evidence

Device product

InCore Lapidus Disposable Kit 28mm Right, Part Number: IC-LAP-28R - Product Usage: Intended as a three-part construct intended for internal fixation for First Metatarsocuneiform arthrodesis.

Z-1501-2021

March 26, 2021

Class II

Product summary

Firm
Nextremity Solutions
Event
Event 87665
Status
Terminated
Classification
Class II
Quantity
123 units
Official record key
device-enforcement:Z-1501-2021

Official wording

Reason: Product could potentially contain an undersized inner diameter of the 40mm drill bushing which result in not allowing the drill bit to pass through the drill bushing, causing inability of surgeon to complete surgical procedure as intended

Code information: Lot Number: 168B29619B UDI: 00817701020875

Distribution pattern: AL, AK, CA, IA,IL, IN, KY ,MD. MI, NJ, NY, OR, PA, SC, TX, VA

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Product could potentially contain an undersized inner diameter of the 40mm drill bushing which result in not allowing the drill bit to pass through the drill bushing, causing inability of surgeon to complete surgical procedure as intended