Device product
AQUABEAM Handpiece, REF: HP2000, a sterile single-use component of the AQUABEAM Robotic System.
Z-2330-2021
Product summary
- Event
- Event 88337
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 608
- Official record key
device-enforcement:Z-2330-2021
Official wording
Reason: Scope tube tip may detach from the telescoping tube, which is attached to a single-use handpiece, a component of an aquablation robotic system. Both tip detachment and entire tip separation form the handpiece may lead to procedural delay for handpiece exchange. Tip detachment can be seen cystoscopically and tip separation from the handpiece may necessitate tip removal from the bladder or prostate.
Code information: Lot: 21C00226; 21C00290; 21C00304; 21C00359; 21C00373; 21C00464; 21C00465; 21C00527
Distribution pattern: Worldwide distribution - US Nationwide distribution in the states of FL, IL, TX, WI, AZ, AR, MI, WV, SD, PA, CA, DC, NJ, NY, GA, IN, VA, TN and the countries of Austria, Switzerland, Germany, Spain, France, UK, Italy.
Derived failure modes
-
Unknown
Scope tube tip may detach from the telescoping tube, which is attached to a single-use handpiece, a component of an aquablation robotic system. Both tip detachment and entire tip separation form the handpiece may lead to procedural delay for handpiece exchange. Tip detachment can be seen cystoscopically and tip separation from the handpiece may necessitate tip removal from the bladder or prostate.