Device product

Naturalyte Liquid Acid concentrate bottles, Product Code 08-2301-3 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

Z-0835-2021

November 23, 2020

Class II

Product summary

Firm
Fresenius Medical Care Holdings, Inc.
Event
Event 86939
Status
Ongoing
Classification
Class II
Quantity
1426 cases
Official record key
device-enforcement:Z-0835-2021

Official wording

Reason: The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.

Code information: Lot Numbers: 20NXAC013 20NXAC035 Additional Affected lots as of 3/17/21: 20LXAC056 20LXAC041 20LXAC055 20LXAC057 20NXAC013 20NXAC035

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.