Device product
BioPlex 2200 ANA Screen Control Set, REF: 663-1131 incompatible with ANA Screen Reagent Packs/ANA Kits, REF: 665-1150; and ANA Screen Calibrator Sets, REF: 663-1101
Z-0082-2022
Product summary
- Event
- Event 88652
- Status
- Terminated
- Classification
- Class II
- Quantity
- 206
- Official record key
device-enforcement:Z-0082-2022
Official wording
Reason: A new lot of ANA screen control and any subsequent control lots are incompatible with older lot numbers of ANA calibrators and ANA reagent packs.
Code information: REF: 663-1131, Lot: 51765, Expiration: 03/05/2021, and any subsequent lots incompatible with: REF: 665-1150, Lot-Expiration: 300884-11/30/2019, 300907-1/31/2020, 300928-3/31/2020, 300933-4/30/2020; and REF: 663-1101, Lot-Expiration: 50661-11/30/2019, 50726-3/11/2020, 50835-3/27/2020, 50858-5/1/2020
Distribution pattern: Worldwide distribution - U.S Nationwide distribution including in the states of CA, FL, NJ, IL, NE, TN, IN, TX, LA, WA, MI, MA, MN, OH, HI, MO, PA, CT, UT, SD, NY, AL, PR, OR, MD, KY, NC, AZ, SC, VA, WI, GA, IA, RI, AK, OK, KS, ND, MS, ME, NM, AR and the countries of ES, AE, DE, FR, KR, CN, TW, JP, IL, SG, AU, NZ, HK, IT, CA, PT, GB, NO, CZ, SE, CH, UA, RU, BE, DK.
Derived failure modes
-
Unknown
A new lot of ANA screen control and any subsequent control lots are incompatible with older lot numbers of ANA calibrators and ANA reagent packs.