Device product

1688 Camera Control Unit (CCU), Catalog number:1688010000. Used in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery wherever a laparoscope/endoscope/arthroscope/sinuscope is used.

Z-0728-2022

January 14, 2022

Class II

Product summary

Firm
Stryker Corporation
Event
Event 89515
Status
Ongoing
Classification
Class II
Quantity
11,955
Official record key
device-enforcement:Z-0728-2022

Official wording

Reason: A software defect in the camera control unit (CCU) will cause the image on the monitor to flip upside-down into an incorrect orientation. The potential of harms are conversion to open procedure, additional medical intervention, or a revision surgery.

Code information: All serial numbers of 1688 CCU, UDI: 07613327420081, with Software Revisions: 3.0.6, 4.0.13, and 4.0.18

Distribution pattern: Worldwide distribution. US nationwide including Puerto Rico. OUS: BR, AE, NL, IN, FR, ES, HK, IT, GB, CO, PE, MY, TH, CA, JP, MX, AR, NP, SG, PH, ZA, CL, KR, CN

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software defect