Device product
S-Series Equipment Management System; Model Number(s):0682400550; UDI: 07613327400502 Components that are affected within the S-Series Finished Good Model Number o P38198 - ASM, MFR CONTROLLER, FIXED STANDARD o P38199 - ASM, MFR CONTROLLER, LIFT STANDARD o P40553 - ASM, MFR HANDLE REPAIR, LIFTING, S-SERIES, STANDARD o P40554 - ASM, MFR HANDLE REPAIR, FIXED, S-SERIES, STANDARD o P46269 - ASSY, MFR CONTROLLER UPGRADE, STANDARD
Z-1297-2021
Product summary
- Event
- Event 87331
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 1,393 units
- Official record key
device-enforcement:Z-1297-2021
Official wording
Reason: The manufacturer has become aware that there is a likelihood that the S-Series Equipment Management System Capacitive Touch Controls on the MFRs rotational electrical magnetic brakes may disengage when an electrocautery device is in use.
Code information: Serial Numbers: 2258046 2863490
Distribution pattern: Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV, and WY. The country of Canada.
Derived failure modes
-
Unknown
The manufacturer has become aware that there is a likelihood that the S-Series Equipment Management System Capacitive Touch Controls on the MFRs rotational electrical magnetic brakes may disengage when an electrocautery device is in use.