Device product

Ambu VivaSight 2 DLT; endobronchial tube system used for verifying tube placement and repositioning. Catalog numbers 412351000, 412371000, 412391000, 412411000

Z-1420-2022

June 03, 2022

Class II

Product summary

Firm
Ambu Inc.
Event
Event 90488
Status
Ongoing
Classification
Class II
Quantity
12,921
Official record key
device-enforcement:Z-1420-2022

Official wording

Reason: Firm has received complaints on the device concerning leak/rupture of either the bronchial or tracheal cuff. Leakage during procedure could lead to loss of secure airway and will require re-intubation of patient.

Code information: Catalog No: 412351000 UDI-DI: 5707480145706 1 each 5707480145713 5 pc/pack Catalog No. 412371000 UDI-DI: 5707480145737 1 each 5707480145744 5 pc/pack Catalog No. 412391000 UDI-DI: 5707480145768 1 each 5707480145775 5 pc/pack Catalog No. 412411000 UDI-DI: 5707480145799 1 each 5707480145805 5 pc/pack All lots

Distribution pattern: Domestic distribution nationwide. Foreign distribution to Austria, Belgium, Cyprus, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Luxembourg, Netherlands, Norge, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, UK Canada, Australia, New Zealand, Israel .

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Firm has received complaints on the device concerning leak/rupture of either the bronchial or tracheal cuff. Leakage during procedure could lead to loss of secure airway and will require re-intubation of patient.