Device product
Transseptal Needle with Catheter. RPN TSN-17-75.0-ENDRYS. GPN G19261. Coaxial set consisting of an outer catheter, curved-tip metal cannula, and a tapered-tip inner needle
Z-0223-2022
Product summary
- Firm
- Cook Inc.
- Event
- Event 88808
- Status
- Terminated
- Classification
- Class II
- Quantity
- 4789 (OUS only)
- Official record key
device-enforcement:Z-0223-2022
Official wording
Reason: Transseptal needles may contain rust on the interior and/or exterior of the needle, which may not be detected by the user. Potential adverse events if an affected product is used ranges from increased procedural time to inflammatory reactions, which at their most severe include systemic reactions potentially leading to permanent impairment or death.
Code information: All Lots UDI (01)00827002192614(17)211010(10)9226180 (01)00827002192614(17)211015(10)9234901 (01)00827002192614(17)211015(10)9235946 (01)00827002192614(17)211015(10)9235947 (01)00827002192614(17)211017(10)9242172 (01)00827002192614(17)211120(10)9326411 (01)00827002192614(17)211120(10)9326412 (01)00827002192614(17)211121(10)9328622 (01)00827002192614(17)211130(10)9347132 (01)00827002192614(17)211203(10)9351027 (01)00827002192614(17)211204(10)9354000 (01)00827002192614(17)211210(10)9370544 (01)00827002192614(17)211220(10)9396895 (01)00827002192614(17)220104(10)9419991 (01)00827002192614(17)220109(10)9431068 (01)00827002192614(17)220109(10)9431069 (01)00827002192614(17)220124(10)9472321 (01)00827002192614(17)220128(10)9480776 (01)00827002192614(17)220201(10)9492814 (01)00827002192614(17)220201(10)9492815 (01)00827002192614(17)220201(10)9492816 (01)00827002192614(17)220214(10)9524330 (01)00827002192614(17)220222(10)9545708 (01)00827002192614(17)220222(10)9545720 (01)00827002192614(17)220304(10)9568083 (01)00827002192614(17)220304(10)9568088 (01)00827002192614(17)220311(10)9583267 (01)00827002192614(17)220319(10)9604183 (01)00827002192614(17)220319(10)9604184 (01)00827002192614(17)220322(10)9611406 (01)00827002192614(17)220322(10)9613905 (01)00827002192614(17)220325(10)9617635 (01)00827002192614(17)220328(10)9626175 (01)00827002192614(17)220328(10)9626176 (01)00827002192614(17)220328(10)9626177 (01)00827002192614(17)220402(10)9638768 (01)00827002192614(17)220404(10)9643348 (01)00827002192614(17)220404(10)9643349 (01)00827002192614(17)220409(10)9654385 (01)00827002192614(17)220409(10)9654513 (01)00827002192614(17)220409(10)9654514 (01)00827002192614(17)220409(10)9655757 (01)00827002192614(17)220510(10)9730630 (01)00827002192614(17)220516(10)9743413 (01)00827002192614(17)220516(10)9743414 (01)00827002192614(17)220522(10)9760371 (01)00827002192614(17)220611(10)9802149 (01)00827002192614(17)220620(10)9826728 (01)00827002192614(17)220620(10)9826729 (01)00827002192614(17)220627(10)9840896 (01)00827002192614(17)220628(10)9844564 (01)00827002192614(17)220701(10)9850898 (01)00827002192614(17)220822(10)9966055 (01)00827002192614(17)220826(10)9971573 (01)00827002192614(17)220826(10)9971744 (01)00827002192614(17)220910(10)10002117 (01)00827002192614(17)220917(10)10016709 (01)00827002192614(17)220919(10)10025480 (01)00827002192614(17)220923(10)10031245 (01)00827002192614(17)221023(10)10102218 (01)00827002192614(17)221023(10)10102219 (01)00827002192614(17)221024(10)10105936 (01)00827002192614(17)221024(10)10105937 (01)00827002192614(17)221113(10)10148679 (01)00827002192614(17)221114(10)10151314 (01)00827002192614(17)221118(10)10157368 (01)00827002192614(17)221121(10)10166249 (01)00827002192614(17)221219(10)10224967 (01)00827002192614(17)221219(10)10224968 (01)00827002192614(17)230204(10)10307597 (01)00827002192614(17)230212(10)10324494 (01)00827002192614(17)230220(10)10342526 (01)00827002192614(17)230220(10)10342529 (01)00827002192614(17)230421(10)13145971 (01)00827002192614(17)230421(10)13145972 (01)00827002192614(17)230514(10)13193150 (01)00827002192614(17)230514(10)13193151 (01)00827002192614(17)230609(10)13235106 (01)00827002192614(17)230609(10)13235107 (01)00827002192614(17)230622(10)13260302 (01)00827002192614(17)230714(10)13304618 (01)00827002192614(17)230716(10)13311621 (01)00827002192614(17)230716(10)13311622 (01)00827002192614(17)230716(10)13311623 (01)00827002192614(17)231009(10)13481714 (01)00827002192614(17)231009(10)13481715 (01)00827002192614(17)231112(10)13553814 (01)00827002192614(17)231202(10)13602045 (01)00827002192614(17)231202(10)13602046 (01)00827002192614(17)231202(10)13602047 (01)00827002192614(17)231204(10)13606368 (01)00827002192614(17)231204(10)13606369 (01)00827002192614(17)231204(10)13606370 (01)00827002192614(17)231204(10)13606371 (01)00827002192614(17)240120(10)13702632 (01)00827002192614(17)240120(10)13702633 (01)00827002192614(17)240120(10)13702634 (01)00827002192614(17)240122(10)13707721 (01)00827002192614(17)240311(10)13817370 (01)00827002192614(17)240311(10)13817371 (01)00827002192614(17)240311(10)13817372 (01)00827002192614(17)220222(10)9545708X (01)00827002192614(17)191110(10)NS7433921 (01)00827002192614(17)191201(10)NS7482244 (01)00827002192614(17)191201(10)NS7482246 (01)00827002192614(17)191201(10)NS7482247 (01)00827002192614(17)200712(10)NS8059999 (01)00827002192614(17)211003(10)NS9209497 (01)00827002192614(17)211003(10)NS9209498 (01)00827002192614(17)211003(10)NS9209499 (01)00827002192614(17)211003(10)NS9209500 (01)00827002192614(17)211004(10)NS9212134
Distribution pattern: Domestic distribution nationwide. Foreign distribution worldwide.
Derived failure modes
-
Unknown
Transseptal needles may contain rust on the interior and/or exterior of the needle, which may not be detected by the user. Potential adverse events if an affected product is used ranges from increased procedural time to inflammatory reactions, which at their most severe include systemic reactions potentially leading to permanent impairment or death.