Skip to content
Recall Observatory FDA recall evidence

Device product

Transseptal Needle with Catheter. RPN TSN-17-75.0-ENDRYS. GPN G19261. Coaxial set consisting of an outer catheter, curved-tip metal cannula, and a tapered-tip inner needle

Z-0223-2022

October 08, 2021

Class II

Product summary

Firm
Cook Inc.
Event
Event 88808
Status
Terminated
Classification
Class II
Quantity
4789 (OUS only)
Official record key
device-enforcement:Z-0223-2022

Official wording

Reason: Transseptal needles may contain rust on the interior and/or exterior of the needle, which may not be detected by the user. Potential adverse events if an affected product is used ranges from increased procedural time to inflammatory reactions, which at their most severe include systemic reactions potentially leading to permanent impairment or death.

Code information: All Lots UDI (01)00827002192614(17)211010(10)9226180 (01)00827002192614(17)211015(10)9234901 (01)00827002192614(17)211015(10)9235946 (01)00827002192614(17)211015(10)9235947 (01)00827002192614(17)211017(10)9242172 (01)00827002192614(17)211120(10)9326411 (01)00827002192614(17)211120(10)9326412 (01)00827002192614(17)211121(10)9328622 (01)00827002192614(17)211130(10)9347132 (01)00827002192614(17)211203(10)9351027 (01)00827002192614(17)211204(10)9354000 (01)00827002192614(17)211210(10)9370544 (01)00827002192614(17)211220(10)9396895 (01)00827002192614(17)220104(10)9419991 (01)00827002192614(17)220109(10)9431068 (01)00827002192614(17)220109(10)9431069 (01)00827002192614(17)220124(10)9472321 (01)00827002192614(17)220128(10)9480776 (01)00827002192614(17)220201(10)9492814 (01)00827002192614(17)220201(10)9492815 (01)00827002192614(17)220201(10)9492816 (01)00827002192614(17)220214(10)9524330 (01)00827002192614(17)220222(10)9545708 (01)00827002192614(17)220222(10)9545720 (01)00827002192614(17)220304(10)9568083 (01)00827002192614(17)220304(10)9568088 (01)00827002192614(17)220311(10)9583267 (01)00827002192614(17)220319(10)9604183 (01)00827002192614(17)220319(10)9604184 (01)00827002192614(17)220322(10)9611406 (01)00827002192614(17)220322(10)9613905 (01)00827002192614(17)220325(10)9617635 (01)00827002192614(17)220328(10)9626175 (01)00827002192614(17)220328(10)9626176 (01)00827002192614(17)220328(10)9626177 (01)00827002192614(17)220402(10)9638768 (01)00827002192614(17)220404(10)9643348 (01)00827002192614(17)220404(10)9643349 (01)00827002192614(17)220409(10)9654385 (01)00827002192614(17)220409(10)9654513 (01)00827002192614(17)220409(10)9654514 (01)00827002192614(17)220409(10)9655757 (01)00827002192614(17)220510(10)9730630 (01)00827002192614(17)220516(10)9743413 (01)00827002192614(17)220516(10)9743414 (01)00827002192614(17)220522(10)9760371 (01)00827002192614(17)220611(10)9802149 (01)00827002192614(17)220620(10)9826728 (01)00827002192614(17)220620(10)9826729 (01)00827002192614(17)220627(10)9840896 (01)00827002192614(17)220628(10)9844564 (01)00827002192614(17)220701(10)9850898 (01)00827002192614(17)220822(10)9966055 (01)00827002192614(17)220826(10)9971573 (01)00827002192614(17)220826(10)9971744 (01)00827002192614(17)220910(10)10002117 (01)00827002192614(17)220917(10)10016709 (01)00827002192614(17)220919(10)10025480 (01)00827002192614(17)220923(10)10031245 (01)00827002192614(17)221023(10)10102218 (01)00827002192614(17)221023(10)10102219 (01)00827002192614(17)221024(10)10105936 (01)00827002192614(17)221024(10)10105937 (01)00827002192614(17)221113(10)10148679 (01)00827002192614(17)221114(10)10151314 (01)00827002192614(17)221118(10)10157368 (01)00827002192614(17)221121(10)10166249 (01)00827002192614(17)221219(10)10224967 (01)00827002192614(17)221219(10)10224968 (01)00827002192614(17)230204(10)10307597 (01)00827002192614(17)230212(10)10324494 (01)00827002192614(17)230220(10)10342526 (01)00827002192614(17)230220(10)10342529 (01)00827002192614(17)230421(10)13145971 (01)00827002192614(17)230421(10)13145972 (01)00827002192614(17)230514(10)13193150 (01)00827002192614(17)230514(10)13193151 (01)00827002192614(17)230609(10)13235106 (01)00827002192614(17)230609(10)13235107 (01)00827002192614(17)230622(10)13260302 (01)00827002192614(17)230714(10)13304618 (01)00827002192614(17)230716(10)13311621 (01)00827002192614(17)230716(10)13311622 (01)00827002192614(17)230716(10)13311623 (01)00827002192614(17)231009(10)13481714 (01)00827002192614(17)231009(10)13481715 (01)00827002192614(17)231112(10)13553814 (01)00827002192614(17)231202(10)13602045 (01)00827002192614(17)231202(10)13602046 (01)00827002192614(17)231202(10)13602047 (01)00827002192614(17)231204(10)13606368 (01)00827002192614(17)231204(10)13606369 (01)00827002192614(17)231204(10)13606370 (01)00827002192614(17)231204(10)13606371 (01)00827002192614(17)240120(10)13702632 (01)00827002192614(17)240120(10)13702633 (01)00827002192614(17)240120(10)13702634 (01)00827002192614(17)240122(10)13707721 (01)00827002192614(17)240311(10)13817370 (01)00827002192614(17)240311(10)13817371 (01)00827002192614(17)240311(10)13817372 (01)00827002192614(17)220222(10)9545708X (01)00827002192614(17)191110(10)NS7433921 (01)00827002192614(17)191201(10)NS7482244 (01)00827002192614(17)191201(10)NS7482246 (01)00827002192614(17)191201(10)NS7482247 (01)00827002192614(17)200712(10)NS8059999 (01)00827002192614(17)211003(10)NS9209497 (01)00827002192614(17)211003(10)NS9209498 (01)00827002192614(17)211003(10)NS9209499 (01)00827002192614(17)211003(10)NS9209500 (01)00827002192614(17)211004(10)NS9212134

Distribution pattern: Domestic distribution nationwide. Foreign distribution worldwide.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Transseptal needles may contain rust on the interior and/or exterior of the needle, which may not be detected by the user. Potential adverse events if an affected product is used ranges from increased procedural time to inflammatory reactions, which at their most severe include systemic reactions potentially leading to permanent impairment or death.