Device product
Panther Fusion Extraction Reagent-X, REF PRD-04477, For use with Panther Fusion System, IVD, Rx only, Contents: FCR-X Fusion Capture reagent-X 4 EA, FER-X Fusion Enhancer Reagent-X (Lithium Hydroxide Solution) 4 EA
Z-0750-2022
Product summary
- Firm
- Hologic, Inc
- Event
- Event 89533
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 264 kits
- Official record key
device-enforcement:Z-0750-2022
Official wording
Reason: Failed stability specifications for reagent kits that may result in false negative test results.
Code information: Panther Fusion Extraction Reagent-X - Lot Numbers: 281375 and 291829 Components included within the kit: Panther Fusion Capture Reagent-X (FCR-X) - Lot Numbers: 281330 and 290312 Panther Fusion Enhancer Reagent-X (FER-X) - Lot Numbers281332 and 290331
Distribution pattern: Worldwide distribution - US Nationwide distribution in the states of AR, CA, CT, FL, GA, ID, IL, IN, MA, MD, ME, MI, NJ, NV, NY, OH, OR, PA, TN, TX, UT, VA, WA, WI, WV, District of Columbia and the countries of Canada, Australia, Belgium, Netherlands, China, Italy, Germany, Sweden.
Derived failure modes
-
Potency or specification failure
Failed stability specifications