Device product

GE Healthcare CARESCAPE Central Station (CSCS) V2.0x MAI700 (Integrated), REF 2082278-001, and MAS700 (Desktop, REF 2082279-001.

Z-0808-2022

January 28, 2022

Class II

Product summary

Firm
GE Healthcare, LLC
Event
Event 89576
Status
Ongoing
Classification
Class II
Quantity
10,445 units
Official record key
device-enforcement:Z-0808-2022

Official wording

Reason: If the CARESCAPE Central Station v2.0 is used with an unapproved keyboard, the audio can be muted resulting in loss of audible alarms.

Code information: Version 2.0.x (v2.0.0, v2.0.1, v2.0.2, or v2.0.3) with serial numbers beginning with SKN for the MAI700 Integrated, UDI 00840682109666 and serial numbers beginning with SNF for the MAI700 Integrated, UDI 0100840682109604; or serial numbers beginning with SNF for the MAS700 Desktop, UDI 00840682109604.

Distribution pattern: Worldwide distribution - US nationwide, including Puerto Rico. There was also government/military distribution. The countries of Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Colombia, Costa Rica, Cote d'lvoire, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, Germany, Ghana, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Republic of Korea, Kuwait, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Mexico, Myanmar, Netherlands, Nigeria, Norway, Pakistan, Palestine, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, Senegal, Singapore, Slovakia, South Africa, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uganda, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    If the CARESCAPE Central Station v2.0 is used with an unapproved keyboard, the audio can be muted resulting in loss of audible alarms.