Skip to content
Recall Observatory FDA recall evidence

Device product

t:slim X2 insulin pump; t:slim X2 insulin pump with Dexcom G5; t:slim X2 insulin pump with Basal-IQ technology; t: slim x2 insulin pump with the Control-IQ Technology

Z-1812-2022

May 24, 2022

Class II

Product summary

Firm
Tandem Diabetes Care Inc
Event
Event 90628
Status
Ongoing
Classification
Class II
Quantity
8399
Official record key
device-enforcement:Z-1812-2022

Official wording

Reason: Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.

Code information: Insulin Pump/Software Versions: t:slim X2 insulin pump/6.4.1 and earlier; t:slim X2 insulin pump with Dexcom G5/5.2.1, 5.2.2; t:slim X2 insulin pump with Basal-IQ technology/6.3.0.1, 6.4, 6.4.1; t: slim x2 insulin pump with the Control-IQ Technology/7.4, 7.4.3

Distribution pattern: US: NC, OK, ME, NY, TX, LA, NE, FL, AZ, IL, OH, CA, AL, MI, UT, VA, MA, SC, MS, GA, NV, HI, MN, WA, IN, AR, ID, CO, IA, MO, MT, PA, WI, NH, MD, WV, KS, NM, OR, KY, DE, ND, CT, NJ, TN, WY, AK, SD, DC, RI, IA, PR, VT, PT, VI, NC. OUS: Australia, New Zealand, South Africa

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software update. Issues could result in