CombiDiagnost R90 Software Version R1.0 and R1.1

Z-2357-2021

July 22, 2020

Class II

Product summary

Firm
Philips Medical Systems Gmbh, DMC
Event
Event 88393
Status
Ongoing
Classification
Class II
Quantity
322 systems; 66 in US
Official record key
device-enforcement:Z-2357-2021

Official wording

Reason: Philips has become aware of a software issue, where the kV and mA values are not locked immediately but change when the collimator shutter position is changed directly after the Lock-In command. This software issue leads to an over or under exposed image.

Code information: Model 706100

Distribution pattern: US Nationwide Distribution

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software issue