Device product
Heater Cooler Unit (HCU 30) The device supplies temperature-controlled water for cardioplegia heat exchangers and for blood heat exchangers in extracorporeal circulation and for blankets with which patients can be warmed or cooled.
Z-2140-2021
Product summary
- Event
- Event 88285
- Status
- Terminated
- Classification
- Class II
- Quantity
- 33
- Official record key
device-enforcement:Z-2140-2021
Official wording
Reason: There is the potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. There is a potential risk of bacterial contamination including Mycobacteria species.
Code information: HCU 30 Base Unit 100-120 V, Model Number 704629 I 70103.4371 HCU 30 Base Unit Canadian/US, Model Number 705501 I 70103.4653HCU 30 Base Unit 200-240 V, Model Number 939001 I 70102.8718 All serial numbers are affected.
Distribution pattern: Distribution to US states of AL, CA, FL, IL, KS, MI, MO, NC, NE & OH.
Derived failure modes
-
Microbial contamination
bacterial contamination