Skip to content
Recall Observatory FDA recall evidence

Device product

FULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST2 220V, Model #317-07927GE CHINA

Z-0370-2023

August 31, 2022

Class II

Product summary

Firm
Full Vision Inc
Event
Event 90995
Status
Ongoing
Classification
Class II
Quantity
8 units
Official record key
device-enforcement:Z-0370-2023

Official wording

Reason: This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.

Code information: UDI/DI 00860176000675, Serial Numbers: GEDC-6765, GEDC-6766, GEDC-6767, GEDC-6768, GEDC-6849, GEDC-6850, GEDC-6851, GEDC-6852

Distribution pattern: Wi, GA, NY, Mexico

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.