Device product

Monaco RTP System, Software Builds 5.40.00, 5.40.01, 5.40.02, and 5.51.10

Z-2406-2021

July 26, 2021

Class II

Product summary

Firm
Elekta Inc
Event
Event 88451
Status
Ongoing
Classification
Class II
Quantity
N/A
Official record key
device-enforcement:Z-2406-2021

Official wording

Reason: If a couch is removed while a plan is loaded, and then added back to the Studyset when the plan is unloaded, it is possible to calculate dose with the couch unassigned to the beams.

Code information: Software Builds: 5.40.00, 5.40.01, 5.40.02, 5.51.10 UDI: (01)00858164002190(10) 5.40.00 (01)00858164002190(10) 5.40.01 (01)00858164002190(10) 5.40.02 (01)00858164002268(10) 5.51.10

Distribution pattern: Distribution in US - IA, NJ, NY, TN, PA, TX, and WI OUS distribution in Australia, Bahrain, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Japan, Korea, Netherlands, Spain, Sweden, Switzerland, Thailand, Turkey and United Kingdom

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    If a couch is removed while a plan is loaded, and then added back to the Studyset when the plan is unloaded, it is possible to calculate dose with the couch unassigned to the beams.