Device product
Simpleware ScanIP software. For transfer of imaging information from a medical scanner to an output file, and for simulating/evaluating surgical treatment options.
Z-2253-2021
Product summary
- Event
- Event 88207
- Status
- Terminated
- Classification
- Class II
- Quantity
- 4 softwares
- Official record key
device-enforcement:Z-2253-2021
Official wording
Reason: A issue (bug) has been identified with the interface and image software which could result in anatomical orientation tags/labels being displayed incorrectly. This could result in operator misunderstanding and errors during resampling or re-alignment of image data
Code information: Affected version/UDI Simpleware ScanIP 7.0: Created before UDI-DI system (released December 2nd, 2014), Simpleware ScanIP 2016.09: 00863520000307, Simpleware ScanIP M-2017.06: 00863520000314, Simpleware ScanIP N-2018.03: 00863520000321
Distribution pattern: US: Nationwide Distribution in the states of CA, FL and countries of : France, United Kingdom, Germany, Australia
Derived failure modes
-
Device software or design
issue (bug) has been identified with the interface and image software