GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Part Numbers 00024307504, 00024330004, 00024345004
Z-0357-2024
Product summary
- Event
- Event 93313
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 2448 total
- Official record key
device-enforcement:Z-0357-2024
Official wording
Reason: Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.
Code information: Part No. 00024307504: UDI-DI: 08426950087922, Lot Numbers: 324230M, 323630G; Part No. 00024330004: UDI-DI: 08426950082507, Lot Numbers: 325655C, 322355C, 323755H, 323855E, 326255D; Part No. 00024345004: UDI-DI: 08426950082538, Lot Numbers: 325660C, 324160J, 323760H
Distribution pattern: US States: MA, MN, NY. India.
Derived failure modes
-
Unknown
Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.