Skip to content
Recall Observatory FDA recall evidence

Device product

Cardinal Health Suction Canister Liner, MEDI-VAC CRD 1000 mL Suction Canister Liner Kit with 6 mm 1.8 m Tube, REF 65651-517

Z-0756-2024

December 21, 2023

Class II

Product summary

Firm
Cardinal Health 200, LLC
Event
Event 93628
Status
Ongoing
Classification
Class II
Quantity
74515
Official record key
device-enforcement:Z-0756-2024

Official wording

Reason: The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister.

Code information: 50630140152670 (case), Lot Numbers: J308-625, J308-629, J308-637, J309-616, J309-634, J309-646, J309-647, J310-603, J310-618

Distribution pattern: US, Canada, Australia and New Zealand

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister.