Device product
Cardinal Health Suction Canister Liner, MEDI-VAC CRD 1000 mL Suction Canister Liner Kit with 6 mm 1.8 m Tube, REF 65651-517
Z-0756-2024
Product summary
- Event
- Event 93628
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 74515
- Official record key
device-enforcement:Z-0756-2024
Official wording
Reason: The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister.
Code information: 50630140152670 (case), Lot Numbers: J308-625, J308-629, J308-637, J309-616, J309-634, J309-646, J309-647, J310-603, J310-618
Distribution pattern: US, Canada, Australia and New Zealand
Derived failure modes
-
Unknown
The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister.