Device product
Hillrom Hospital Bed Accessory, Watch Care Incontinence Management System for Centrella Bed, Product Code/Part Number P00697905
Z-0142-2023
Product summary
- Event
- Event 90865
- Status
- Ongoing
- Classification
- Class I
- Quantity
- 814 units
- Official record key
device-enforcement:Z-0142-2023
Official wording
Reason: Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).
Code information: UDI/DI 887761984998, all serial numbers
Distribution pattern: US
Derived failure modes
-
Unknown
Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).