Philips Allura Xper FD series with Software Version Number: 2.1.x- intended for Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. Model Number: 722005, 722006, 722008, 722010, 722012, 722013, 722025, 722026, 722027, 722028.

Z-0416-2022

November 19, 2021

Class II

Product summary

Firm
Philips North America Llc
Event
Event 89203
Status
Terminated
Classification
Class II
Quantity
USA: 5 systems R.O.W.: 15 systems
Official record key
device-enforcement:Z-0416-2022

Official wording

Reason: Set screws securing the two gliding rods of the Extra Monitors (called 7th and 8th monitor) mounted on the Monitor Ceiling Suspension (MCS) can come loose, the Extra monitors might disengage from the Mounting Ceiling Suspension structure and fall down, may lead to patient, user or service engineer harm that may require medical intervention.

Code information: US Serial Number/UDI: 1 (01)00884838054202(21)1; 527 (01)00884838059054(21)527; 408 (01)00884838054202(21)408; 306 (01)00884838054202(21)306; 61 (01)00884838059030(21)61 OUS: Model Product S/N UDI: 722006 Allura Xper FD20 1997 (01)00884838059191(10)1997; 722010 Allura Xper FD10 1177 (01)00884838059030(21)1177; 722010 Allura Xper FD10 1178 (01)00884838059030(21)1178; 722013 Allura Xper FD20 Biplane 544;; 722010 Allura Xper FD10 514 (01)00884838059030(21)514; 722013 Allura Xper FD20 Biplane 271; 722025 Allura Xper FD20 Biplane OR Table 6; 722005 Allura Xper FD10/10 519; 722008 Allura Xper FD20 Biplane 531; 722012 Allura Xper FD20 1901 (01)00884838059054(21)1901; 722012 Allura Xper FD20 2505 (01)00884838059054(21)2505; 722013 Allura Xper FD20 Biplane 137; 722026 Allura Xper FD10 439 (01)00884838054189(21)439; 722134 Field ext. Xper vascular systems R7.6 98736; 722027 Allura Xper FD10/10 172 (01)00884838054196(21)172;

Distribution pattern: Worldwide distribution - US Nationwide distribution in the states of DC, GA, MA, TX, VA and the countries of Dominican Republic, INDIA, Japan, Poland, Spain, Taiwan.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Set screws securing the two gliding rods of the Extra Monitors (called 7th and 8th monitor) mounted on the Monitor Ceiling Suspension (MCS) can come loose, the Extra monitors might disengage from the Mounting Ceiling Suspension structure and fall down, may lead to patient, user or service engineer harm that may require medical intervention.