Skip to content
Recall Observatory FDA recall evidence

Device product

Omnipod 5 App (on compatible Android smartphones), Software Versions 1.1 - 1.2.3 (current). US: PT-000599- Omnipod 5 App Software Package (provided through Google Play), EU: PT-001323- Omnipod 5 App Software Package (provided on locked-down Controllers (PT-001194 and PT-001193). Used to control an insulin pump.

Z-0591-2024

November 30, 2023

Class I

Product summary

Firm
Insulet Corporation
Event
Event 93511
Status
Ongoing
Classification
Class I
Quantity
27000 users
Official record key
device-enforcement:Z-0591-2024

Official wording

Reason: The bolus calculator is not recording the decimal point if it is the first value entered when changing a bolus dose. This may lead to an over-delivery of insulin to the user if the user does not recognize the error on the bolus calculator screen or the confirmation screen prior to starting the bolus.

Code information: UDI/Model/NDC Number: " Omnipod 5 App Software UDI: 1038508112030 " EU locked-down Controller UDI: 10385083000190 and 10385083000206

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Germany and UK.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The bolus calculator is not recording the decimal point if it is the first value entered when changing a bolus dose. This may lead to an over-delivery of insulin to the user if the user does not recognize the error on the bolus calculator screen or the confirmation screen prior to starting the bolus.