Device product
PCF-H190L, EVIS EXERA III COLONOVIDEOSCOPE
Z-0419-2024
Product summary
- Event
- Event 93335
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 24 incorrectly repaired units
- Official record key
device-enforcement:Z-0419-2024
Official wording
Reason: Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
Code information: UDI-DI 04953170305191 Serial Numbers 2302719, 2516751
Distribution pattern: Domestic distribution to FL, KY, MA, NJ, NY, and PA
Derived failure modes
-
Unknown
Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.