Device product

Triton Canister System (finished part numbers FG 12009, for 3L canisters, and FG 12019 for 2L canisters) Scanning Label (component part number 11032). A software application used to help estimation of blood loss. Used with blood bearing canisters and accessories.

Z-1402-2022

January 22, 2020

Class II

Product summary

Firm
Stryker Instruments Div. of Stryker Corporation
Event
Event 90464
Status
Terminated
Classification
Class II
Quantity
16,700 units
Official record key
device-enforcement:Z-1402-2022

Official wording

Reason: Some Canister Scanning Labels used to calibrate the Triton Canister System have a color variance. Based on a comprehensive analysis, blood loss estimates obtained using these Canister Scanning Labels may result in lower than expected results.

Code information: UDI-DI GTIN: 10859506006101, 10859506006125 All Canister Scanning Labels with a date of manufacture prior to to 2019-12-24

Distribution pattern: US Nationwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Some Canister Scanning Labels used to calibrate the Triton Canister System have a color variance. Based on a comprehensive analysis, blood loss estimates obtained using these Canister Scanning Labels may result in lower than expected results.