Skip to content
Recall Observatory FDA recall evidence

Device product

BD Kiestra InoqulA+ with BD Kiestra" InoqulA+" and BeA Software version 5.1 and 5.1.1 as part of the Icefall A platform. An in vitro diagnostic medical device intended to automate specimen processing according to user-defined procedures and protocols. Catalog (Ref) No. : 447213

Z-1415-2023

March 22, 2023

Class II

Product summary

Firm
Becton Dickinson & Co.
Event
Event 91937
Status
Ongoing
Classification
Class II
Quantity
7 instruments
Official record key
device-enforcement:Z-1415-2023

Official wording

Reason: Upon installation of BD Kiestra" InoqulA+" and BeA version 5.1 and 5.1.1 as part of the Icefall A platform, plate information was not visible in Synapsys" after processing has occurred. This could possibly lead to a delayed diagnosis and/or treatment or need for specimen recollection streaking plated media, and inoculating tubes and slides. In Semi-Automated (SA) mode, plates are automatically selected, barcoded, streaked in a pre-configured pattern while the user manually inoculates plates, tubes, and slides. An optional biosafety cabinet on the SA module provides personnel, product, and environmental protection. The InoqulA+" solution is indicated for use in the clinical laboratory.

Code information: UDI: 038290LSUVLPKFPV Serial Numbers: INO-000156, INO-000249, INO-000248, INO-000252, INO-000254, 20130276, 19033ST002

Distribution pattern: Worldwide - US Nationwide distribution in the states of IL, NY and the countries of Netherlands, Saudi Arabia.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Upon installation of BD Kiestra" InoqulA+" and BeA version 5.1 and 5.1.1 as part of the Icefall A platform, plate information was not visible in Synapsys" after processing has occurred. This could possibly lead to a delayed diagnosis and/or treatment or need for specimen recollection streaking plated media, and inoculating tubes and slides. In Semi-Automated (SA) mode, plates are automatically selected, barcoded, streaked in a pre-configured pattern while the user manually inoculates plates, tubes, and slides. An optional biosafety cabinet on the SA module provides personnel, product, and environmental protection. The InoqulA+" solution is indicated for use in the clinical laboratory.