Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, Custom Packs, REF BB811. Used in Cardiopulmonary bypass.

Z-1023-2023

December 12, 2022

Class II

Product summary

Firm
Medtronic Perfusion Systems
Event
Event 91343
Status
Ongoing
Classification
Class II
Quantity
2,937 units
Official record key
device-enforcement:Z-1023-2023

Official wording

Reason: An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.

Code information: Model BB811; Serial Number Range: 8111483548 - 8113999999; GTIN Number: 00643169178168 (Lot Numbers: 13386860, 13386998, 13387454, 223834103, 224773022, 224773023, 225467084, 225467085); GTIN Number: 00643169178175 (Lot Number: 13388339, 13389098, 13394289, 223815345, 223829391, 223834104, 223834105, 224602715, 224695594, 224702830, 224711400, 224773018, 224773019, 224773020, 224773021, 224875617, 224883237, 224893902)

Distribution pattern: Worldwide distribution. US Nationwide, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Belgium, Plurinational State of Bolivia, Bosnia And Herzegovina, Botswana, Brazil, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Ethiopia, Finland, France, Germany, Ghana, Greece, Guatemala, Honduras, Hungary, Iceland, Indonesia, Islamic Republic of Iran, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Kyrgyzstan, Latvia, Lithuania, Luxembourg, Malaysia, Mexico, Namibia, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Oman, Pakistan, Paraguay, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Slovenia, South Africa, Spain, Suriname, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Turkey, Turkmenistan, Uganda, United Arab Emirates, United Kingdom, Uruguay, Viet Nam, and Yemen.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.